Installing, conditioning and maintaining columns
Maximum operating temperatures
Maximum operating temperatures for stationary phases are usually quoted assuming isothermal operation with a flame ionisation detector. Other detectors may impose different limits, the mass spectrometer being much more susceptible to bleeding of the stationary phase than the thermal conductivity detector. All phases bleed very slightly at high temperatures because of the loss of smaller-sized (and hence lower-boiling) polymer chains, although normally this is not noticeable. Operating temperature has a profound effect on column life, particularly for capillary columns. Loss of stationary phase, or breakdown of the thin film into pools to expose part of the tubing surface, results in serious loss of performance. Additionally, in columns that contain PSX phases with two different functional groups, one group (usually that which confers additional polarity) is preferentially lost. This results in a change of relative separation as well as a loss of resolution. The temperature limit of a column may be determined by the deactivation procedure used in production, rather than by the stationary phase itself. Newer silica columns have a very low metal oxide content, thought to act as a catalyst for the degradation of both sample and stationary phase, and thus enable phases to be run at higher temperatures. Fused silica capillary columns have a protective external coating of polyimide that is slowly degraded at elevated temperatures (maximum temperature 360C, but now to 400C), which can also limit column life. However, separations are usually achieved at much lower temperatures.
0
0
الاكثر قراءة في تحليل السموم
اخر الاخبار
اخبار العتبة العباسية المقدسة