Installing, conditioning and maintaining columns Guard columns and retention gaps
المؤلف:
Sue Jickells , Adam Negrusz (Editors)
المصدر:
Clarkes Analytical Forensic Toxicology
الجزء والصفحة:
480
2026-11-21
41
Installing, conditioning and maintaining columns
Guard columns and retention gaps
A guard column and a retention gap are essentially the same thing, but are installed to serve different purposes. These 1–10 m lengths of fused silica tubing are attached to the front of the chromatography column via a press-snap connector or zero dead-volume union, and then installed into the injector port. The surface of the silica is deactivated to minimise solute inter actions, but no stationary phase is added. The tubing diameter should be the same as that of the column, but if different it should, ideally, be of a wider bore.
The function of a guard column is to trap deposits of nonvolatile residues, thereby preventing their contamination of the analytical column. Solutes are not retained by the guard column (since there is no stationary phase) and pass directly onto the column. Portions can be cut periodically from the top of the guard column as deterioration in chromatography requires, without any appreciable loss of resolution from the analytical column. A retention gap is used to improve peak shape either when poor chromatography is the result of a large injection volume (>2 μL) or when there are solvent–stationary phase polarity mismatches. Greatest improvement is seen in early-eluting peaks, or for solutes with similar polarity to that of the solvent.