Chromatographic principles
The retention of a drug with a given packing material and eluent can be expressed as a retention time or retention volume, but both of these are dependent on flow rate, column length and column diameter. The retention is best described as a column capacity ratio (k), which is independent of these factors. The column capacity ratio of a compound (A) is defined by equation (19.1):

where VA is the elution volume of A and Vm (in some texts denoted by V0) is the elution volume of a non-retained compound (i.e. void volume). At constant flow rate, retention times (tA and tm, referred to in some texts as t0) can be used instead of retention volumes. The injection of a solvent or salt solution can be used to measure V0, but the solute used should always be recorded along with reported k data. The importance of selecting suitable solutes for the measurement of V0 has been discussed (Wells and Clark 1981). It is sometimes convenient to express retention data relative to a known internal standard (B). The ratio of retention times (tA /tB) can be used, but the ratio of adjusted retention times, (tA-t0)/ (tB-t0), is better when data need to be transferred between different chromatographs (Ettre 1980).
Resolution is the parameter that describes the separation power of the complete chromato graphic system relative to the particular compo nents of the mixture. By convention, resolution (R) is expressed as the ratio of the distance between two peak maxima (VR,2-VR,1) to the mean value of the peak width ([W1+W2]/2) at the baseline (equation 19.2):

If we approximate peaks by symmetric tri angles, then if R is equal to or more than 1.5, the components are completely separated. If R is less than 1.5, the components overlap. Sensitivity in chromatographic analysis is a measure of the smallest detectable level of a component in a chromatographic separation and is dependent on the signal-to-noise ratio in a given detector. Sensitivity can be increased by derivatisation of the compound of interest, optimisation of chromatographic system or miniaturization of the system. The limit of detection is normally taken as 3 times the signal-to-noise (S/N) ratio and the limit of quantification as 10 times the S/N ratio.