Sample preparation and sample presentation
Sampling techniques for Raman spectroscopy are relatively simple, since the only requirements are that the monochromatic laser beam irradiates the sample of interest and the scattered radiation is focused upon the detector.
Raman spectroscopy may be performed on very small samples (e.g. a few nanograms). Powders do not need to be pressed into disks, the material simply has to be irradiated by the laser beam. Solid samples are often examined in gold coated or glass sample holders that generally require about 25–50 mg of material. Typically, liquid samples are analysed in quartz or glass cuvettes, which may have mirrored rear surfaces to improve the signal intensity. Glass is a very weak Raman scatterer and so many samples (liquid and solid) can be analysed directly in a bottle or, for example, in a nuclear magnetic resonance (NMR) tube, although fluorescence from some types of glass can be problematic.
Water is a good solvent for Raman studies, since the Raman spectrum of water is essentially one broad, weak band at 3500 cm1. One sample type that may pose a problem is darkly coloured material. Often, these samples absorb excessive heat and burn, which causes sample degradation and spectral degradation. Amorphous materials also have a tendency to absorb heat in the laser beam. To avoid sample burning, it may be necessary to dissipate the heat. This can be accomplished by reducing the laser power or by using an accessory that spins the sample and so avoids irradiation of a single point in the sample. Further reduction of the laser power can also be accomplished through a neutral density filter. Sometimes, the sample can be diluted in KBr to help reduce sample burning.
The complete Stokes Raman spectrum, which covers shifts in the range 100–3500 cm1, can be obtained and the intensity of Raman scattering is directly proportional to the concentration of the scattering species, an important factor for quantitative analysis. However, the Raman effect is relatively weak and hence the material needs to be present at a concentration of at least about 0.1–0.5% for accurate assessments, whereas IR can be used to detect materials down to a concentration of approximately 0.01%. As noted above, fluorescence can also be problematic in Raman studies, but typically arises from additives in the glass sample tubes or impurities within the sample of interest. Data massaging techniques can sometimes blank out Raman spectral contributions from fluorescent materials. Photobleaching is another way to suppress fluorescence. This technique involves irradiating the sample for a prolonged period of time (seconds to hours) with the laser. During this time the fluorescence may decrease through the destruction of the fluorescing component from the prolonged exposure to the laser irradiation. The spectrum is then acquired after photo bleaching is complete. Of course, it is also possible that the laser radiation may change the component of interest. For example, a solvated crystalline material may desolvate with exposure to laser radiation. It is wise to establish sample integrity by comparing spectra acquired with short acquisition times to those acquired with long acquisition times. Additionally, another technique, such as IR spectroscopy or X-ray powder diffraction, could be used to check for sample integrity after Raman analysis. Variable-temperature studies in Raman spectroscopy provide a wealth of information. A Raman spectrum typically covers a wavelength range that extends beyond the range normally associated with mid-IR spectroscopy (typically 4000–400 cm-1), so information about the lattice vibrations of compounds is readily available. By varying the temperature of a sample, the lattice energies of the compound can be changed, which allows the nature of the crystal lattice to be interpreted. In addition, information similar to that obtained in IR variable-temperature studies (crystal form changes and the nature of solvate association) can be obtained with variable-temperature Raman investigations.