Thin-layer chromatography
Suitable systems for thin-layer chromatography (TLC) . After separation, solutes are present on the TLC plate as spots or bands intimately associated with the stationary phase. TLC separations can be carried out by normal-phase TLC or by reversed-phase operation. Normal-phase TLC typically employs silica gel as the stationary phase. Reversed-phase TLC uses organic stationary phases, typically long-chain hydrocar bons, covalently bonded to silica gel. Reversed phase TLC should be avoided if IR is to be used for identification because it is difficult to remove the spot without also removing the organic stationary phase, giving rise to a mixed IR spec trum. Furthermore, location reagents must be chosen with care, and a destructive reagent, such as the Marquis reagent for alkaloids (which contains concentrated sulfuric acid), should not be used. Nondestructive reagents such as iodoplatinate solution can be used because the coloured complex is decomposable to yield the original compound. However, even this procedure may introduce extraneous peaks into the spectrum and, ideally, location reagents are best avoided. If the compound cannot be detected under UV light, it could be applied to the thin layer plate twice and only a portion of the chromatogram sprayed, which thus allows location of substances but also permits the unsprayed portion to be eluted.
The use of aqueous acid or alkali to elute the compound from the thin-layer plate, followed by solvent extraction of the aqueous solution, is more efficient than direct solvent extraction of the adsorbent. In one method of direct extraction, the adsorbent is scraped from around the spot, the glass adjacent to the spot is carefully cleaned and the adsorbent is eluted in situ directly onto a wall of potassium bromide (KBr) built around the tip of the spot. The KBr is then pressed into a disc. This technique is only suitable for well-resolved spots. Elution of the spot sideways reduces contamination from compounds that are not resolved as well. The recovery of material from chromatograms varies from nil to over 70%. Compounds that contain hydroxyl and carboxyl groups, which can readily form hydrogen bonds with the solid support, tend to be recovered in low yield. Considerable interference in the 1100 cm-1 region is found with some adsorbents and compounds.
In a variation of this method, the thin-layer adsorbent is placed in the bottom of a glass vessel together with a triangular ‘wick’ of compressed KBr. Solvent is added and it rises up the wick and evaporates from the upper region. The compound is conveyed up the wick by the solvent and accumulates at the tip of the triangle, which is then cut off, dried, and used to prepare a disc. About 10 lg of compound is required to produce a satisfactory spectrum. The advantage of this technique is that the lower part of the KBr wick acts as a filter and removes finely divided adsorbent, which can give rise to spurious peaks.
In a further method, the thin-layer adsorbent is scraped onto a small amount of KBr powder in the hub of an 18-gauge metal hypodermic needle. A 1 mL glass syringe is filled with pure solvent and connected to the needle, and the compound is eluted dropwise onto a mound (10 mg) of dry KBr powder. Each drop of solvent is allowed to evaporate completely before the elution of the next drop. The powder and solute are then mixed and pressed into a disc.
Eluted material almost always includes unwanted extraneous matter co-extracted from the thin-layer chromatogram. Thus, it is advisable to use the eluent from a ‘blank’ area as a reference solution. Contamination from plasticisers, solvents and dirty glassware can also be a serious problem when a spectrum has to be obtained from a few micrograms of a compound. Even momentary contact of dry adsorbent with plastic tubing can remove appreciable quantities of plasticisers. Hence the following precautions should be taken:
• use the minimum amount of the purest adsorbent available
• elute with less than 1 mL of a solvent that contains <0.0001% (1 ppm) of nonvolatile residue
• keep sample handling to a minimum
• clean all glassware with an efficient detergent in an ultrasonic bath
• avoid contact of materials and samples with plastics.